Inficon T-Guard Leak Detection Sensor Interface descriptio Manual do Utilizador Página 12

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 40
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 11
12 RS232 ASCII Interface Protocol
jins85e1 2.fm Interface Description(1101)
Command
word1
Command
word2
Command
word3
Parameter Statement Read/
Set
Example Response
*CONFig :TIME :AUTo <b> automatic setting of times
b: enable = 1; disable = 0 S
R
*CONF:TIME:AUT 1
*CONF:TIME:AUT?
OK
ENABLED
:AUToPurge automatic purge in accumulation mode
<b> b: enable = 1; disable = 0 S
R
*CONF:TIME:AUTP 1
*CONF:TIME:AUTP?
OK
ENABLED
:MEASure total measurement time
<value> value = 0 .. 300 sec S
R
*CONF:TIME:MEAS 23.3
*CONF:TIME:MEAS?
OK
23.3
:PURGE purge time in accumulation mode
<value> value = 1.. 50 sec S
R
*CONF:TIME:PURGE 1
*CONF:TIME:PURGE?
OK
1.0
:WaitPurge waiting time before automatic purge in
accumulation mode
<value> value = 0 .. 300 sec S
R
*CONF:TIME:WP 1
*CONF:TIME:WP?
OK
1.0
Vista de página 11
1 2 ... 7 8 9 10 11 12 13 14 15 16 17 ... 39 40

Comentários a estes Manuais

Sem comentários