Inficon T-Guard Leak Detection Sensor Interface descriptio Manual do Utilizador Página 20

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 40
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 19
20 RS232 ASCII Interface Protocol
jins85e1 2.fm Interface Description(1101)
2.4.2 Carrier Gas Mode
2.4.2.1 Performing a calibration
Every command must be terminated by the two bytes
Carriage Return (CR, ASCII 13, \r) and Line Feed (LF, ASCII 10, \n), in this order.
The response you receive from T-Guard will also be terminated like this.
Command sent to
T-Guard
Response received
from HLD5000
Content
*STAT:MEAS? READY status of the unit
*CONF:MODE CARGAS OK switch to carrier gas mode
*CAL:STAT? NO CAL RUNNING no calibration running
*CONF:TRIG2ON OFF OK turn off trigger 2
*CONF:TIME:AUT ON OK turn on automatic times
*CONF:UNIT:FU SCCM OK set flow unit to sccm
*CONF:CF 30000 OK set carrier gas flow to 30000
*CONF:TRIG1:MBAR*L/S
1E-3
OK set trigger 1 to 1E-3 mbar*l/s
*CONF:TIME:MEAS? 5.0 read measurement time
*CONF:TLR:MBAR*L/S 8E-4 OK set test leak rate to 8E-4 mbar*l/s
*CAL:START OK start calibration measurement
*CAL:STAT? CAL RUNNING, WAIT calibration running
*CAL:STAT? CAL FINISHED, CONFIRM calibration finished
*STAT:ERR? NO ERROR/WARNING no errors
*CAL:FAC:OLD? 1.017 read old calibration factor
*CAL:FAC:NEW? 1.098 read new calibration factor
*CAL:QUIT OK confirm calibration
Vista de página 19
1 2 ... 15 16 17 18 19 20 21 22 23 24 25 ... 39 40

Comentários a estes Manuais

Sem comentários