Inficon T-Guard Leak Detection Sensor Manual do Utilizador Página 54

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 85
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 53
52 Operation
jina85e 5.fm Technical Handbook(1112)
5.6 Configuring analog output
You can configure the analog output in the menu screen
Main Menu
Settings
Interfaces
in Analog Output and Analog Scale
In the setting "LR Exponent", the exponent results from the voltage as follows:
Exponent = (voltage x 2) -14
For example: A voltage of 4.5 V results in an exponent of (4.5 x 2) - 14 = -5.
In the setting "LR Log.", the leak rate is calculated as follows:
Leak rate = final value x (10^((U - 10)/(volts/decade)) )
For example: With a final value of 10
-1
mbar l/s and 2 V per decade, a voltage of 3.7
V results in a leak rate of 10
-1
mbar l/s x 10^((3.7-10)/2) = 10
-1
mbar l/s x 10
-3,15
=
10
-1
mbar l/s x 7,08x10
-4
= 7,08x10
-5
mbar l/s.
For "Current Log.", it is similar for current instead of leak rate.
Fig. 6-14 Configuration of analog outputs.
Fig. 6-15 Configuration of the scale of the analog outputs.
Vista de página 53
1 2 ... 49 50 51 52 53 54 55 56 57 58 59 ... 84 85

Comentários a estes Manuais

Sem comentários